Veterinary Use CONTEC CA10M VET Mainstream ETCO2 Capnograph Respiration Rate End-tidal CO2 Monitor For Animals
$413.49
Description
Details:
Introduction
The device adopts mainstream method to measure end-tidal carbon dioxide(EtCO2) and airway respiratory rate(AwRR). It can monitor the pet’s ventilation function, reflect the condition of pulmonary circulation and pulmonary blood flow, and indirectly reflect the alveolar ventilation. It has important application value in clinical anesthesia, cardiopulmonary cerebral resuscitation, prehospital emergency care, ICU and post-anesthesia of pets. It connects with the pet’s respiratory tract through the air adapter and respiratory connection tube.
Features
1)End-tidal CO2 measurement function;
2)Respiration rate measurement function;
3)Optional three units;
4)With audible and visual alarm and silencing functions, the alarm limit can be set;
5)Over-limit alarm function for End-tidal CO2 and respiration rate;
6)Low battery alarm function;
7)With the prompt function of ‘Adapter off’;
8)Users can calibrate zero by themselves.
9)With atmospheric pressure and temperature compensation functions;
10)Waterproof and dustproof grade grade:IP22;
11)Degree of protection against electric shock: type BF applied part
12)Install 2 AAA dry batteries, up to 4-hour user time.
13)It can be used in highland enviroments.
Performance
End-tidal CO2 measurement:
1)Measurement range: 0 ~150mmHg
2)Resolution: 1mmHg
3)Accuracy:
0~40 mmHg: ±2 mmHg
41~70 mmHg: ±5% of readings
71~100 mmHg: ±8% of readings
101~150 mmHg: ±10% of readings
Respiration rate measurement:
1)Measurement range: 2rpm~150rpm
2)Resolution: 1rpm
3)Accuracy: ±1rpm
Safety characteristic:
1)Device type: internally powered equipment
2)Degree of protection against electric shock: type BF applied part
3)Protection grade: IP22
Other parameters:
1)Preheat time: 15s
2)Response time: < 3s
3)Alarm sound pressure level: 50 dB
4)Battery life time: 4hours
Accessories
★ A User Manual
★ An air adapter
★ 2 AAA dry batteries
Physical Characteristic
Dimension: 59×56×59.5mm
Weight: 92g (with batteries)
Working environment
Temperature: 5℃~40℃
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 3V (2 AAA dry batteries)
Input power: ≤1VA
Additional information
MPN | D0100H7CDW2 |
---|---|
CONDITION | New |
MODEL | CA10MVET |
Reviews
There are no reviews yet.